6482-901-01 Rev. A / August 2012
Return to
8-15
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 8: Triggering
Figure 8-9
Remote operation of the trigger model
Event detection
Once the instrument is taken out of an idle state, operation proceeds through the trigger model to
perform the source, delay, and measure actions.
Operation through the trigger model is delayed at the event detector until the programmed trigger
event occurs. If an event detector has a bypass (:DIRection), operation can be programmed to
loop around the event detector.
Arm-in
event
Arm event
detector
Trigger delay
SOURCE
action
Delay event
detector
DELAY
action
Measure
event
detector
MEASURE
action
No
Yes
Idle
Another
arm
?
Yes
No
Arm
layer
= GPIB default
Trigger-in
source
1
1
0.0 sec
No
See Note
?
Yes
SOURce
ARM
:DIRection
ACCeptor
TRIGger
:DIRection
SOURce
ACCeptor
TRIGger:INPut
SOURce
TRIGger:DELay
<n>
SOURce
TRIGger:OUTPut*
DELay
TRIGger:OUTPut*
TRIGger:OUTPut*
SENSe
TRIGger:INPut
TRIGger:INPut
SENSe
DELay
TRIGger:COUNt
<n>
ARM:OUTPut
NONE|TRIGger
ARM:COUNt
<n>|INF
*
GPIB default parameter for TRIGGER:OUTPut is NONE
**
SOURceX = SOURce[1] or SOURce2
+
+
+
+
+
+
+
+
Note:
The following commands
place the Model 6482 into an
idle state: DCL, SDC, ABORt,
*RST,
SYSTem:PREset,
and*RCL
Trigger
layer
TRIGger:SOURce
+ IMMediate
TLINk
= Output trigger
SOURceX:DELay
<n>|AUTO**
+
0.0005 sec
Another
trigger
?
INITiate
Source event
detector
ARM :SOURce
+ IMMediate
BUS
TIMer
MANual
TLINk
NStest
PStest
BStest
Summary of Contents for 6482