7-10
Return to
6482-901-01 Rev. A / August 2012
Section 7: Sweep Operation
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
NOTE
6. From the CONFIGURE SWEEPS menu, select
SWEEP-COUNT
, press the
ENTER
key, and
then select
FINITE
or
INFINITE
.
7. From the CONFIGURE SWEEPS menu, select
SOURCE-RANGING
, press the
ENTER
key,
and then select
BEST-FIXED
,
AUTO-RANGE
, or
FIXED
.
8. Press the
EXIT
key to return to the normal display.
Step 3. Set the delay.
Set the source delay as follows:
1. Press the
CONFIG
key, and then depending on the selected source channel, press the
SRC1
or
SRC2
key.
2. Select
DELAY
, and then press the
ENTER
key.
3. Set the delay value, and then press the
ENTER
key.
4. Press the
EXIT
key to return to the normal display.
Step 4. Turn the output on.
Press the
ON/OFF OUTPUT
key to turn the output on (blue OUTPUT indicator turns on). The
Model 6482 will output the programmed bias level.
Step 5. Run the sweep.
To run the sweep, press the
SWEEP
key. After the sweep is completed, turn the output off by
pressing the
ON/OFF OUTPUT
key.
Summary of Contents for 6482
Page 3: ......