Appendix F: Example Programs
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Introduction
This section provides listings of pseudo-code for example programs throughout this manual.
These programs include:
•
Basic measurement program from
.
•
.
•
Sweep program from
.
•
Limit test program from
Basic measurement program
The program listing below performs basic single-channel device under test (DUT) measurement
as covered in
. This program sets up the following operating modes:
•
Channel 2 measurement range: 2
μ
A
•
Channel 2 source range: 10 V
•
Channel 2 source output level: 10 V
'Basic Measurement Program
*RST
' Restore GPIB Defaults
:SENS2:CURR:RANG 2E-6
' Select 2 uA Channel 2 measure range
:FORM:ELEM CURR2
' Select Channel 2 reading.
:SOUR2:VOLT:RANG 10
' Select 10 V Channel 2 source range.
:SOUR2:VOLT 10
' Output 10 V on channel 2.
:OUTP2 ON
' Turn on channel 2 output.
:READ?
' Trigger and request reading.
' Read back result
:OUTP2 OFF
' Turn off channel 2 output.