![background image](http://html1.mh-extra.com/html/keithley/6482/6482_reference-manual_4059393255.webp)
16-62
Return to
6482-901-01 Rev. A / August 2012
Section 16: SCPI Command Reference
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Description
This command is used to select input lines for the Trigger Link. For normal
operation, Trigger Link input and output (“OLINe”) should not share the same line.
OLINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OLINe <NRf>Select output line; arm
layer
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe <NRf>Select output line; trigger
layer
Parameters
<NRf> =
1
Line #1
2
Line #2
3
Line #3
4
Line #4
5
Line #5
6
Line #6
Query
:OLINe?
Query output trigger line
Description
This command is used to select output lines for the Trigger Link. For normal
operation, Trigger Link input and output (“ILINe”) should not share the same line.
OUTPut <event list>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OUTPut <event list>Arm layer event
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut <event list>Trigger layer events
Parameters
Arm Layer Trigger
<event list> = TRIGger
Trigger on exiting trigger layer
NONE
Disable arm layer output trigger
Trigger Layer Triggers
<event list> = SOURce
Output trigger after source level is set
DELay
Output trigger after delay period
SENSe
Output Trigger after measurement
NONE
Disable trigger layer triggers
NOTE
Each event in the list must be separated by a comma (for example, :arm:output source,
delay, sense).
Query
:OUTPut?
Query output trigger event(s)
Description
This command is used to specify when trigger pulses occur on the specified
output trigger line of the Trigger Link (:OLINe).
Arm Layer Trigger:
With TRIGger selected, an output trigger will occur when
exiting the trigger layer. With NONE selected, the arm layer output trigger is dis-
abled.
Trigger Layer Triggers
: You can specify from one to all three events. Each event
in the list must be separated by a comma (,).
The SOURce, DELay and MEASure events refer to the source-delay-measure
(SDM) cycle. This is the Device Action in the Trigger Model. See
and
. With SOURce specified, an output trigger occurs after the source is
set. With DELay specified, an output trigger occurs after the delay period. With
MEASure specified, an output trigger occurs after the measurement.
NOTE
When disabling triggers, the NONE parameter must be sent alone (for example, trig:outp
none). If it is listed with any of the other event parameters, NONE will be ignored.
Summary of Contents for 6482
Page 3: ......