Appendix B: Calibration
Model 2657A High Power System SourceMeter® Instrument Reference Manual
B-6
2657A-901-01 Rev. B/December 2012
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range is
coupled to the source range, and you cannot independently control the measure range. Thus, it is not
necessary for you to set the range when testing voltage or current measurement accuracy.
Current source accuracy
Follow the steps below to verify that the Model 2657A output current accuracy is within specified
limits.
Figure 126: Connections for current verification
LINE RATING
100-240 VAC
50, 60 Hz
350 VA MAX.
LINE FUSE
SLOWBLOW
5.0 A, 250 V
NO INTERNAL OPERATOR SERVICEABLE PARTS,
SERVICE BY QUALIFIED PERSONNEL ONLY.
REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
INPUT/OUTPUT CURRENT: 120 mA MAX.
RS-232
DIGITAL I/O
LAN
!
IEEE-488
!
!
SENSE
LO
HI
LO
GUARD
SENSE
HI
LO
3030 VDC
MAX.
250 V
MAX.
LO
HI
Input
LO
Model 2657A
Amps
Test fixture
Digital multimeter
Pin 22 = +5 V
Pin 24 = Interlock (INT)