Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-286
2657A-901-01 Rev. B/December 2012
Bit Value
Description
B0-B9 Not
used
Not applicable
B10
status.operation.instrument.trigger_timer.TRIGGER_OVERRUN
status.operation.instrument.trigger_timer.TRGOVR
Set bit indicates one
or more enabled bits
for the operation
status trigger timer
overrun register is set.
Bit B10 decimal value:
1,024
Binary value:
0100 0000 0000
B11-B15 Not
used
Not applicable
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the bit to set.
For example, to set bit B10, set
operationRegister
to 1024.
Example
status.operation.instrument.trigger_timer.enable = 1024
Uses a decimal value to set
the TRGOVR bit of the
operation status trigger timer
summary enable register.
Also see
status.operation.instrument.trigger_timer.trigger_overrun.*
(on page 7-287)