Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-136
2657A-901-01 Rev. B/December 2012
localnode.model
This attribute stores the model number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
model
= localnode.model
model
The model number of the instrument
Details
When using this command from a remote node,
localnode
should be replaced with the node reference, for
example
node[5].model
.
Example
print(localnode.model)
Outputs the model number.
Also see
localnode.password
This attribute stores the remote access password.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (W)
Yes
LAN reset
LAN restore defaults
Nonvolatile memory
""
Usage
localnode.password =
passWord
passWord
String containing the remote interface password
Details
This write-only attribute stores the password that is set for any remote interface. When password usage is
enabled (
localnode.passwordmode
), you must supply this password to change the configuration or to control
an instrument from a web page or other remote command interface.
The instrument continues to use the old password for all interactions until the command to change it executes.
When changing the password, give the instrument time to execute the command before attempting to use the
new password.
You cannot retrieve a lost password from any command interface.
The password can be reset by resetting the LAN from the front panel or by using the
lan.reset()
command.
When using this command from a remote node,
localnode
should be replaced with the node reference, for
example
node[5].password
.