Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-372
2657A-901-01 Rev. B/December 2012
Usage
waitcomplete()
waitcomplete(
group
)
group
Specifies which TSP-Link group on which to wait
Details
This function will wait for all previously started overlapped commands to complete.
A group number may only be specified when this node is the master node.
If no
group
is specified, the local group is used.
If zero (0) is specified for the
group
, this function waits for all nodes in the system.
Any nodes that are not assigned to a group (group number is 0) are part of the master node's group.
Example 1
waitcomplete()
Waits for all nodes in the local group.
Example 2
waitcomplete(G)
Waits for all nodes in group G.
Example 3
waitcomplete(0)
Waits for all nodes on the TSP-Link network.
Also see
None