
Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-344
2657A-901-01 Rev. B/December 2012
Details
This function waits up to
timeout
seconds for a trigger on the active command interface. A command interface
trigger occurs when:
•
A GPIB GET command is detected (GPIB only)
•
A VXI-11 device_trigger method is invoked (VXI-11 only)
•
A *TRG message is received
If one or more of these trigger events were previously detected, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed. This is true
regardless of the number of events detected.
Example
triggered = trigger.wait(10)
print(triggered)
Waits up to 10 seconds for a trigger.
If
false
is returned, no trigger was detected
during the 10-second timeout.
If
true
is returned, a trigger was detected.
Also see
(on page 7-336)
tsplink.group
This attribute is the group number of a TSP-Link node.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Not saved
0
Usage
groupNumber
= tsplink.group
tsplink.group =
groupNumber
groupNumber
The group number of the TSP-Link node (0 to 64)
Details
To remove the node from all groups, set the attribute value to 0.
When the node is turned off, the group number for that node changes to 0.
The master node can be assigned to any group. You can also include other nodes in the group that includes the
master. Note that any nodes that are set to 0 are automatically included in the group that contains the master
node, regardless of the group that is assigned to the master node.
Example
tsplink.group = 3
Assign the instrument to TSP-Link group number 3.
Also see
Using groups to manage nodes on TSP-Link network
(on page 6-57)
Summary of Contents for 2657A
Page 3: ......