Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-345
tsplink.master
This attribute reads the node number assigned to the master node.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
masterNodeNumber
= tsplink.master
masterNodeNumber
The node number of the master node
Details
After doing a TSP-Link reset (
tsplink.reset()
), use this attribute to access the node number of the master in
a set of instruments connected over TSP-Link.
Example
LinkMaster = tsplink.master
Store the TSP-Link master
node number in a variable
called LinkMaster.
Also see
tsplink.node
This attribute defines the node number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile
memory
1
Usage
nodeNumber
= tsplink.node
tsplink.node =
nodeNumber
nodeNumber
Set node to a number (1 to 64)
Details
This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until the next time
tsplink.reset()
is executed on any node
in the system.
Each node connected to the TSP-Link system must be assigned a different node number.
Example
tsplink.node = 2
Sets the TSP-Link node to number 2.
Also see
(on page 7-347)