Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-78
2657A-901-01 Rev. B/December 2012
Usage
func
= display.smu
X
.measure.func
display.smu
X
.measure.func =
func
func
One of the following values:
•
0
or
display.MEASURE_DCAMPS
: Selects current measurement function
•
1
or
display.MEASURE_DCVOLTS
: Selects volts measurement function
•
2
or
display.MEASURE_OHMS
: Selects ohms measurement function
•
3
or
display.MEASURE_WATTS
: Selects power measurement function
X
Source-measure unit (SMU) channel (for example,
display.smua.measure.func
applies to SMU channel A)
Details
Selects the displayed measurement function: Amperes, volts, ohms, or watts.
Example
display.smua.measure.func = display.MEASURE_DCAMPS
Selects the current measure function
for SMU A.
Also see
(on page 7-77)
display.trigger.clear()
This function clears the front-panel trigger event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.trigger.clear()
Details
The trigger event detector remembers if an event has been detected since the last
display.trigger.wait()
call. This function clears the trigger event detector and discards the previous history of TRIG key presses.
This attribute also clears the
display.trigger.overrun
attribute.
Also see
(on page 7-79)
(on page 7-80)