Using TSB for upgrading the firmware
After downloading the new flash file from the Keithley Instruments website, you can use Test Script
Builder (TSB) to upgrade the firmware of your Model 2657A.
1. If not already running, start Test Script Builder (on the PC desktop, double-click the icon for the
Test Script Builder).
2. On the Instrument Console toolbar, click the
Open Instrument
icon and then select your
communication interface from the Select Instrument Resource dialog box. See the section on
TSP Programming Fundamentals for details on opening communications.
Figure 124: Open Instrument icon
3. On the Instrument Console toolbar, click the
Flash memory
icon to display a drop-down menu.
Figure 125: Flash memory icon
4. From the drop-down menu, select
Instrument
and then click
Flash
.
5. From the Select A Firmware Data File dialog box, use the browser to select the file name of the
new firmware and click
Open
to upgrade the firmware.