Section 3: Functions and features
Model 2657A High Power System SourceMeter® Instrument Reference Manual
3-76
2657A-901-01 Rev. B/December 2012
Bit identification for indicators
Bit
B16
B15
B14
B13
B12
B11
B10
B9
Annunciator
REL
REAR
SRQ
LSTN
TALK
REM
ERR
EDIT
Weighted
value* 32768 16384 8192 4096 2048 1024 512 256
Binary
value 0/1 0/1 0/1 0/1 0/1 0/1 0/1 0/1
Bit
B8
B7
B6
B5
B4
B3
B2
B1
Annunciator
SMPL
STAR
TRIG
ARM
AUTO
4W
MATH
FILT
Weighted
value*
128 64 32 16 8 4 2 1
Binary
value 0/1 0/1 0/1 0/1 0/1 0/1 0/1 0/1
* The weighted values are for bits that are set to “1.” Bits set to “0” have no value.
Not all of the above indicators shown in above table may be used by the Model 2657A.
Local lockout
You can use the front-panel EXIT (LOCAL) key to cancel remote operation and return control to the
front panel. However, this key can be locked-out to prevent a test from being interrupted. When
locked, this key becomes a NO-OP (no operation). Configure the following attribute to lock or unlock
the EXIT (LOCAL) key:
display.locallockout =
lockout
Where
lockout
is set to one of the following values:
0
or
display.UNLOCK
1
or
display.LOCK
For example, to lock out the EXIT (LOCAL) key:
display.locallockout = display.LOCK
Load test menu
The LOAD TEST menu lists tests (USER, FACTORY, and SCRIPTS) that can be run from the front
panel. Factory tests are preloaded and saved in nonvolatile memory at the factory. They are available
in the FACTORY TESTS submenu. Named scripts that have been loaded into the run-time
environment can be selected from the SCRIPTS submenu. Refer to
(on page 6-3) for
additional information.
User tests
User tests can be added to or deleted from the USER TESTS submenu.
Summary of Contents for 2657A