Status Structure 8-21
On power-up, all error messages are enabled and will go into the error queue as they occur.
Status messages are not enabled and will not go into the queue. As listed in
, there are
commands to enable and/or disable messages. For these commands, the <list> parameter is used
to specify which messages to enable or disable. The messages are specified by their codes. The
following examples show various forms for using the <list> parameter.
<list>
=
(-110)
Single message
=
(-110:-222)
Range of messages (-110 through -222)
=
(-110:-222, -220)
Range entry and single entry (separated by a comma)
When you enable messages, messages not specified in the list are disabled. When you disable
messages, each listed message is removed from the enabled list.
NOTE
To prevent all messages from entering the error queue, send the enable command
along with the null list parameter as follows: STATus:QUEue:ENABle ().
Table 8-6
SCPI commands — error queue
Command
Description
Default
STATus
:QUEue
[:NEXT]?
:ENABle <list>
:ENABle?
:DISable <list>
:DISable?
:CLEar
STATus subsystem:
Read error queue:
Read and clear oldest error/status message.
Specify error and status messages for error queue.
Read the enabled messages.
Specify messages not to be placed in queue.
Read the disabled messages.
Clear messages from error queue.
(Note 1)
(Note 2)
(Note 2)
SYSTem
:ERRor?
:CLEar
SYSTem subsystem:
Read error queue:
Clear messages from error queue.
(Note 1)
Notes:
1. Power-up and *CLS empties the error queue. STATus:PRESet has no effect.
2. Power-up enables error messages and disables status messages. *CLS and STATus:PRESet have no effect.
Programming example — read error queue
STAT:QUE:ENAB (+000:+900)
‘ Enable all Keithley defined messages (dis-
able all SCPI defined messages).
STAT:QUE?
‘ Return oldest message.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...