8-16
Status Structure
Questionable event status
The used bit of the questionable event register (shown in
) is described as follows:
•
Bit B8, calibration summary (Cal) — Set bit indicates that an invalid calibration
constant was detected during the power-up sequence. This error will clear after
successful calibration of the power supply.
OR
Qusetionable Event
Enable Register
Cal = Calibration Summary
& = Logical AND
OR = Logical OR
&
Decimal
Weights
[:EVENt]?
256
(2
8
)
:CONDition?
Cal
(B8)
(B7-B0)
(B15-B9)
Questionable Condition
Register
Questionable Event
Register
:ENABLe <NRf>
:ENABLe?
To QSB Bit
of Status Byte
Register
Cal
(B8)
(B7-B0)
(B15-B9)
Cal
(B8)
(B7-B0)
(B15-B9)
Figure 8-7
Questionable event status
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...