Performance Verification 13-7
2.
Select the multimeter DC volts measuring function, and enable auto-ranging.
3.
Select Channel #1 by pressing the
▲
and
▲
keys to toggle between Channel #1 and
Channel #2. Make sure the Model 2306 output is turned on.
4.
Verify output voltage accuracy for each of the voltages listed in
. For each test
point:
• Set the Model 2306 output voltage to the indicated value.
• A value of current compliance of 0.25A is sufficient.
• Allow the reading to settle.
• Verify that the multimeter reading is within the limits given in the table.
1.
Repeat steps 1 through 4 for Channel #2 (use the
▲
and
▲
keys to toggle between
Channel #1 and Channel #2).
Table 13-2
Output voltage accuracy limits
Model 2306 output
voltage setting
Output voltage limits
(2 Years, 18°C-28°C)
5.00V
4.9945 to 5.0055V
10.00V
9.992 to 10.008V
15.00V
14.9895 to 15.0105V
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...