Pulse Current Measurements 3-25
Figure 3-5
4 Up
3 Down
0 Up
4 Down
5 Up
0 Down
Sample pulse forms for step method
Figure 3-6
0 Up
5 Down
5 Up
0 Down
Sample one-shot only pulses for step method
Trigger level settings
The trigger level may be set to a unique value for each active step. Use the TLEVx command
to set appropriate trigger levels for each active step in the waveform. Make sure that the maxi-
mum setting for the selected trigger level range is not exceeded. (
has 5 rising edge steps and 4 falling edge steps. Set the trigger levels for each step
measurement according to the expected pulses. Based on the wave form, the nine trigger levels
could be set as follows:
Rising: Falling:
TLEV1
100mA
TLEV6
900mA
TLEV2
300mA
TLEV7
600mA
TLEV3
500mA
TLEV8
400mA
TLEV4
700mA
TLEV9
300mA
TLEV5
900mA
For a programming example of this sample, see
“Sample step method” on page 3-32
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...