6-4
External Triggering (Model 2306-VS Only)
Trigger signals
General waveforms for the trigger signals are shown in
and
. Note that
both input and output triggers may be programmed for either rising-edge or falling-edge
operation (see
topic later in this section).
Figure 6-3
Trigger input signal
Triggers Step On
Leading Edge
High
(3V-5V)
Low
(<0.8V)
2
μ
s
Minimum
A. Falling Edge Input Trigger
Triggers Step On
Leading Edge
High
(3V-5V)
Low
(<0.8V)
2
μ
s
Minimum
B. Rising Edge Input Trigger
Figure 6-4
Trigger output signal
Trigger Occurs at
End of Step
10
μ
s
A. Falling Edge Output Trigger
Triggers Occurs at
End of Step
10
μ
s
B. Rising Edge Output Trigger
>4V
≤
0.8V
>4V
≤
0.8V
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...