10-2
Signal Oriented Measurement Commands
Overview
The signal oriented measurement commands are used to acquire readings. You can use these
high-level instructions to control the measurement process. These commands are summarized
in
Table 10-1
Signal oriented measurement command summary
Command
Description
:FETCh[1]?
Returns the last reading from battery channel (#1). Triggered before
reading(s).
:FETCh[1]:ARRay?
Returns the last array of readings from battery channel (#1). Triggered
before reading(s).
:FETCh2?
Returns the last reading from charger channel (#2). Triggered before
reading(s).
:FETCh2:ARRay?
Returns the last array of readings from charger channel (#2). Triggered
before reading(s).
:READ[1]?
Triggers and returns a new battery channel (#1) reading.
:READ[1]:ARRay?
Triggers and returns a new array of battery channel (#1) readings.
:READ2?
Triggers and returns a new charger channel (#2) reading.
:READ2:ARRay?
Triggers and returns a new array of charger channel (#2) readings.
:MEASure[1][:<function>]?
Performs a READ? on the specified battery channel (#1) function.
:MEASure[1]:ARRay[:<func-
tion>]?
Performs a READ:ARRay? on the specified battery channel (#1) function.
:MEASure2[:<function>]?
Performs a READ2? on the specified charger channel (#2) function.
:MEASure2:ARRay[:<function>]?
Performs a READ2:ARRay? on the specified charger channel (#2)
function.
BOTHTRG
Triggers a reading on channel 1 and then channel 2. After this command
completes, the display is set for Channel #2.
BOTHFETCH?
Responds with channel 1 and channel 2 readings in a single message. The
message contains a value for channel 1, a comma, and then a value for
channel 2. After this command completes, the display is set for Channel
#1.
BOTHREAD?
Triggers reading on channel 1 and then channel 2, then responds with
channel 1 and channel 2 readings in a single message. The message
contains a value for channel 1, a comma, and then a value for channel 2.
After this command completes, the display is set for Channel #2.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...