Table C-1
Remote calibration command summary
Command
Description
:CALibration
Calibration subsystem.
:PROTected
Cal commands protected by password.
:CODE '<code>'
Unlock cal; changes code if cal is already unlocked. (Default
code: KI002306.)
:COUNt?
Query number of times 2306 has been calibrated.
:DATE <yyyy>,<mm>,<dd>
Program calibration year, month, day.
:DATE?
Query calibration year, month, day.
:INIT
Initiate calibration (must be sent before other cal steps).
:SAVE
Save calibration data to EEPROM.*
:LOCK
Lock out calibration. (Abort if calibration is incomplete.)
:STEP0 <nrf>
Output Ch. 1 full-scale voltage (14V).
:STEP1 <nrf>
Calibrate Ch. 1 output voltage using external DMM reading.
:STEP2 <nrf>
Calibrate Ch. 1 voltage measuring using external DMM reading.
:STEP3
Perform Ch. 1 DVM input full-scale (14V) cal.
:STEP4 <nrf>
Output current (1.9A) for Ch. 1 5A full-scale cal.
:STEP5 <nrf>
Calibrate Ch. 1 output current limit using calculated current.
:STEP6 <nrf>
Calibrate Ch. 1 5A measurement range using calculated current.
:STEP7
Output 5mA nominal current for Ch. 1 5mA range full-scale cal.
:STEP8 <nrf>
Calibrate Ch. 1 5mA measurement range.
:STEP9 <nrf>
Output Ch. 2 full-scale voltage (14V).
:STEP10 <nrf>
Calibrate Ch. 2 output voltage setting using DMM reading.
:STEP11 <nrf>
Calibrate Ch. 2 voltage measuring using external DMM reading.
:STEP12
Perform Ch. 2 DVM input full-scale (14V) cal.
:STEP13 <nrf>
Output current (1.9A) for Ch. 2 5A full-scale cal.
:STEP14 <nrf>
Calibrate Ch. 2 output current limit using calculated current.
:STEP15 <nrf>
Calibrate Ch. 2 5A measurement range using calculated current.
:STEP16
Output 5mA nominal current for Ch. 2 5mA range full-scale cal.
:STEP17 <nrf>
Calibrate Ch. 2 5mA measurement range.
Calibration Reference C-3
* Calibration data will not be saved if:
1. Calibration was not unlocked with :CODE command.
2. Invalid data exists. (For example, cal step failed or was aborted.)
3. Incomplete number of cal steps were performed.
4. Calibration was not performed in the proper sequence.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...