3-14
Pulse Current Measurements
:MILLiamp <NRf>
Model 2306-PJ when on 500mA current range. Set trigger level range
(10mA, 100mA, or 500mA). The parameter <NRf> sent with
this command causes the trigger to be set with the trigger level
setting of HALFamp, HUNDred, or TEN. Queries receive
responses of 0.5, 0.1, or 0.01 accordingly. For example, if a
value of 75mA is sent with the command, a value of 0.1A will
be returned as a response to a query.
:DELay <NRf>
Specify trigger delay in seconds:
0.0–0.1 (pulse current measurements) or
0.0–5.0 (pulse current digitization).
0.0
:STEP <b>
Performs a series of measurements (
See “Pulse current step method” on
OFF
:UP
<NRf>
<0-20> (max is for both up and down combined)
1
:DOWN <NRf>
<0-20> (max is for both up and down combined)
1
:TIME <NRf>
33.3
μ
sec–100msec
200
μ
sec
:TimeOUT <NRf>
TimeOUT (other than the first): 2msec–200msec
2ms
:INITial <NRf>
First TimeOUT step: 10msec–60secs
2sec
:DELay <NRf>
0msec–100msec (in 10msec steps)
0
:SEQuence
Define up to 19 delay sequence values using the index and the
time value (e.g. “1,1e-3” sets the first delay to 1ms). Ranges: i: 1-
19 and t: 0 to 100e-3.
0
:LENgth
Define the delay sequence length (0 to 19). Specify zero to
always use the PCUR:STEP:DELay value.
0
:SKIP
Specify whether (1) or not (0) to skip sequential triggers.
1 = Only requires 1 trigger
0 = Requires multiple triggers
:RANGe <NRf>
Model 2306, 2306-VS or 2306-PJ when on 5A current range.
Set trigger level range (100mA, 1A, or 5A). The parameter
<NRf> sent with this command causes the trigger to be set with
the trigger level setting of MILL, ONE, or AMP. Queries receive
responses of 0.1, 1.0, or 5.0 accordingly. In other words, if a
value of 2.0A is sent with the command, a value of 5A will be
returned as a response to a query.
5A
:MILLiamp <NRf>
Model 2306-PJ when on 500mA current range. Set trigger level range
(10mA, 100mA, or 500mA). The parameter <NRf> sent with
this command causes the trigger to be set with the trigger level
setting of HALFamp, HUNDred, or TEN. Queries receive
responses of 0.5, 0.1, or 0.01 accordingly. For example, if a
value of 75mA is sent with the command, a value of 0.1A will
be returned as a response to a query.
:TLEVx <NRf>
Set trigger level for each TLEV step where
x
equals 1–20
(0.0–maxA where max is 100mA for 100mA RANGe setting, 1A
for 1A RANGe setting, and 5A for 5A RANGe setting).
0.0
Table 3-2
SCPI commands — pulse current measurements (cont.)
Command
Description
Default
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...