Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.8 Reprocessing of the patient element
27 / 56
2.8.3 Reprocessing of the spittoon bowl
CAUTION
Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
4
Make sure that the unit is turned off before removing the tumbler filler or
the spittoon bowl.
4
Before turning on the unit, always check if the tumbler filler, tumbler holder
and spittoon bowl are present and in their proper places.
CAUTION
Damage caused by inappropriate cleaning agents.
Damage
4
Do not use acid-containing, sand-containing, and foaming cleaning agents.
4
Add a dose of DEKASEPTOL gel dosing cap to the spittoon bowl and distrib-
ute the agent with a brush.
4
For cleaning, allow the gel to act for approx. 5 minutes.
4
For disinfection, allow the gel to act for approx. 15 minutes.
4
After exposure, press the "Spittoon bowl" button.
4
Clean the spittoon bowl thoroughly using a brush.
Also refer to:
2.9.3 Cleaning programmes "Morning" and "Evening", Page 33
Also refer to:
2.9.5 "Weekly" cleaning programme, Page 35
ð
DEKASEPTOL Gel is dosed and added automatically in the "Evening" and
"Weekly" cleaning programmes.
4
For thorough cleaning or washer disinfection, rotate the spittoon bowl and
lift it off.
4
Check the sieve insert and replace it according to need.