3 Replacing and checking cleaning and germ reduction agents | 3.4 Checking the OXYGENAL 6 concentration
53 / 56
4
Then fill-up the water bottle with tap water.
NOTE
Substances remaining in the dosing head are not contamination, but rather
agents that are deposited there during extended downtimes after condensa-
tion of the water. These deposits dissolve after several dosing processes.
4
Screw the water bottle tightly to the treatment centre in a clockwise rota-
tion.
ð
Apply compressed air to the water bottle.
OXYGENAL 6 volumes for 1.5 litre water bottle
Type of use
Amount of OXYGENAL
6
Doses attachment
Standard concentration
5 ml
1x
Higher concentration
60 ml
12x
3.4 Checking the OXYGENAL 6 concentration
NOTE
Check the OXYGENAL 6 concentration occasionally in order to check the func-
tion of the germ reduction unit for water and/or the expiration date of the
germ reduction agent.