Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.7 Reprocessing the three-function and multifunctional
handpiece
24 / 56
4
Cover the medical device with the KaVo Cleanpac bag, and place it on the
cleaning adapter manual 3F/MF syringe cannula (
Mat. no. 3.005.7780
).
4
Hold the can vertically.
4
Press the spray key for at least 3 seconds.
Also refer to:
Instructions for use KaVo DRYspray
2.7.4 Automated reprocessing
NOTE
The cannula can be subjected to automated reprocessing. Adapters are
needed for automated reprocessing. Order adapter separately. Adapter mach.
3F/MF (
Mat. no. 3.005.9024
)
WARNING
Incomplete disinfection.
Infection hazard.
4
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4
If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
Preparation for automated internal and external
cleaning as well as internal and external disinfection
Miele Series G 7881/7891
Cannula
Requisite material:
Adapter mach. 3F/MF (
Mat. no. 3.005.9024
)
Automated internal and external cleaning and internal
and external disinfection
KaVo recommends washer disinfectors according to EN ISO 15883-1, which are
operated using alkaline cleaning agents having a maximum pH value of 10.
The validation was performed in a Miele washer disinfector using the
"VARIO-TD" programme and the "neodisher MediClean" cleaner.
4
For programme settings as well as cleansers and disinfectants, please refer
to the Instructions for Use of the washer disinfector.
Automated drying
The drying procedure is usually part of the cleaning programme of the washer
disinfector.