Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes
35 / 56
ð
The default value for the cleaning programmes "After treatment" and
"Evening" is 20 seconds per instrument.
ð
The default value for the "Morning" cleaning programme is 2 minutes
per instrument.
NOTE
The default value is based on the current RKI guideline (RKI = German
Robert Koch Institute).
2.9.5 "Weekly" cleaning programme
NOTE
This cleaning programme is not available in combination with the water block.
The "Weekly" cleaning programme prevents the formation of microorganisms
during extended downtimes. During the "Weekly" cleaning programme, the af-
ferent water ducts are automatically filled with a higher concentration of KaVo
OXYGENAL 6. The "Weekly" cleaning programme takes approx. 45 minutes to
complete. The increased hydrogen peroxide concentration acts for at least 30
minutes in the system, or 48 hours over the weekend.
An additional "Weekly" cleaning programme must be run during extended
downtimes (more than 4 weeks). The need for "Weekly" cleaning is displayed
automatically on the display of the dentist element in the "Cleaning" tab at the
initial start-up and after downtimes (weekends, public holidays, vacation etc.):
"Weekly cleaning required. For x days."
NOTE
Observe the instructions for use for the instruments.
NOTE
If the suction hoses are attached in the hygiene centre, the "Evening" clean-
ing programme runs parallel to the "Weekly" cleaning programme.
NOTE
The suction cannot be used while the cleaning programme is in progress.