Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.8 Reprocessing of the patient element
26 / 56
2.8 Reprocessing of the patient element
2.8.1 Reprocessing of the tumbler
4
Reprocess as specified by the manufacturer.
2.8.2 Reprocessing of the tumbler holder and tumbler
filler
CAUTION
Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
4
Make sure that the unit is turned off before removing the tumbler filler or
the spittoon bowl.
4
Before turning on the unit, always check if the tumbler filler, tumbler holder
and spittoon bowl are present and in their proper places.
NOTE
During assembly, make sure that the O-ring is present on the tumbler filler.
Inspect the O-ring (
Mat. no. 0.200.6128
) for damage and exchange it if ne-
cessary to prevent water or oxidation damage.
The tumbler filler and tumbler holder can be subjected to wipe disinfection.
4
Pull the tumbler filler
①
(
Mat. no. 0.764.8541
) upwards and lift off.
4
Take off the tumbler holder
③
(
Mat. no. 0.764.6153
).
4
Clean the tumbler filler and tumbler holder manually.
4
Wet the O-ring
②
(
Mat. no. 0.200.6128
) with silicone grease (
Mat. no.
1.000.6403
) for servicing.
The tumbler filler is thermodisinfectable.
4
Parts should be standing upright when placed in the washer disinfector.