Care instructions KaVo uniQa
3 Replacing and checking cleaning and germ reduction agents | 3.4 Checking the OXYGENAL 6 concentration
54 / 56
4
Test the standard concentration 0.02% H
2
O
2
(200ppm): Fill a measuring
cup (40 ml) with 39 ml water and 1 ml treated water.
4
Immerse a Merckoquant Peroxide Test Strip in the liquid.
ð
Compare the colours and read the value: 5 mg/l is shown for a concentra-
tion of 200ppm.