Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.3 Reprocessing of the surfaces of treatment centre / uphol-
stery / foot control
12 / 56
CAUTION
Product damage due to improper disinfection.
Malfunctions.
4
Use the disinfectant according to the specifications of the manufacturer.
4
No spray disinfection, perform wipe disinfection only.
4
Do not immerse product or parts of the product in liquids.
4
Mop up any spilled cleanser or disinfectant immediately.
4
Clean the surfaces and upholstery with a soft cloth and water, and wipe-dis-
infect them with an approved surface disinfectant.
CAUTION
Damage caused by inappropriate use of water.
Damage to the foot control.
4
For cleaning, the foot control must not be immersed in water or placed un-
der running water.
4
When wiping it off with a moist cloth, do not touch the contacts of the
charge socket.
4
Only wipe off. Do not spray.