![KaVo uniQa Care Instructions Download Page 17](http://html1.mh-extra.com/html/kavo/uniqa/uniqa_care-instructions_4013490017.webp)
Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.5 Reprocessing of the assembly kit for physiological saline
solution and surgical motor
17 / 56
Manual drying
NOTE
Coolant hose/NaCl hose with accessories is intended for single use only and is
not to be disinfected and sterilised. No drying required.
4
Allow all disinfected and sterilised parts to dry fully on room air before using
them again.
2.5.2 Automated reprocessing
CAUTION
Damage by liquids.
Faults on electrical components.
4
Do not subject the device to automated reprocessing.
Automated cleaning and disinfection
The components listed above are released for automated cleaning and disinfec-
tion.
NOTE
Please comply with the corresponding Instructions for use when you repro-
cess the motor and motor cable.
Automated drying
The drying procedure is usually part of the cleaning programme of the washer
disinfector.
NOTE
Please comply with the instructions for use of the washer disinfector.
Service, inspection and testing after preparation
NOTE
It is essential to comply with the hygiene requirements (sterility) during the
test after reprocessing. If any fracturing or obvious surface changes are vis-
ible, the parts need to be checked by the Service.
Check for cleanliness, intactness, servicing, and repair as described in the fol-
lowing:
4
Check the adjustable functions of the unit and the motor function.
4
Check the hose pump for sufficient coolant flow rate.
4
Check the control commands on the foot control.
2.5.3 Packaging
NOTE
The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!
NOTE
If potentially infectious liquids and particles can become deposited on the
products, it is recommended to cover and protect these areas with sterile dis-
posable products.