Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes
30 / 56
4
Grasp the three-function or multifunctional handpiece on the adapter
between hose and handpiece and take off the cannula
②
and the gripping
sleeve
①
.
4
Place the three-function or multifunctional handpieces without cannulas
②
or gripping sleeves
①
in the middle holders of the hygiene centre.
NOTE
Make sure to place three-function handpieces in the grey holder and multi-
functional handpieces in the black holder of the hygiene centre.
4
Place the instrument hoses of the dentist element in the designated places
of the hygiene centre.
4
Place the suction hoses of the assistant element in the designated places of
the hygiene centre.
①
Spray mist suction
④
Three-function handpiece (grey
adapter) or multifunctional hand-
piece (black adapter)
②
Saliva ejector
⑤
With insert: Motor / MULTIflex
coupling
③
Without insert: PiezoLED
NOTE
All water regulation mechanisms on the motor, instruments, and MULTIflex
couplings must be open.