Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.10 Reprocessing of the hygiene centre
44 / 56
ð
The cleaning programme proceeds automatically.
4
After completion of the "Weekly" cleaning programme, place hoses in hold-
ers.
2.10 Reprocessing of the hygiene centre
CAUTION
Damage caused by inappropriate cleaning agents.
Damage
4
Do not use acid-containing, sand-containing, and foaming cleaning agents.
NOTE
The drain must be checked for damage prior to the weekly cleaning pro-
gramme. The handpiece holder of the hygiene centre insert should be cleaned
or disinfected in a washer disinfector unit according to need.
4
Detach and remove the handpiece holder by pressing on the depression.
Summary of Contents for uniQa
Page 1: ...Care instructions KaVo uniQa...
Page 56: ...1 014 7863 Fk 20230117 3 en...