Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes
33 / 56
The "After treatment" cleaning programme prevents proteins from coagulating
and adhering inside the drainage system and improves the removal of fine
particles. In addition, the cleaning programme includes automatic rinsing of the
water-conducting system and all instruments for 20 seconds each.
Tipp: Place the tumbler on the tumbler holder first to prevent the water from
splashing.
4
Run after each treatment.
ð
The following message is displayed after completion of the cleaning pro-
gramme:
The instruments and suction hoses can now be returned to their holders.
2.9.3 Cleaning programmes "Morning" and "Evening"
The "Morning" and "Evening" cleaning programmes are automatic programmes
for hydro-mechanical cleaning and disinfection of the suction hoses, drainage
and separation systems. The cleaning process is carried out by supplying fresh,
clear water and disinfectant. The water is sucked out of the Aquamat through
the extraction point and suction handpiece to the separation system. The spit-
toon bowl rinsing is alternately turned on and off.
The "Morning" cleaning programme includes automatic rinsing of the water-
conducting system and all instruments for 2 minutes each including tumbler
filling. In addition, the suction system is rinsed with water.
In the short cleaning programme "Evening", all instruments on the dentist ele-
ment and assistant element are rinsed for 20 seconds each. In addition, the
suction system and the spittoon bowl are cleaned and disinfected with DEKA-
SEPTOL Gel. KaVo recommends running the "Evening" cleaning programme
after surgical interventions.
Tipp: Place the tumbler on the tumbler holder first to prevent the water from
splashing.
The DEKASEPTOL gel is aspirated automatically. As soon as the DEKASEPTOL
Gel bottle is empty, the display on the dentist element shows a status message.
4
Tap the corresponding button to start the cleaning programme "Morning" or
"Evening".
ð
The following message is displayed after completion of the cleaning pro-
gramme: