6 Reprocessing steps in accordance with ISO 17664 | 6.5 Storage
52 / 68
▪ Steriliser with triple pre-vacuum:
– at least 3 minutes at 134
o
C -1
o
C / +4
o
C (273
o
F -1.6
o
F / +7.4
o
F)
▪ Steriliser using the gravity method:
– at least 10 minutes at 134
o
C -1
o
C / +4
o
C (273
o
F -1.6
o
F / +7.4
o
F)
4
Remove the medical device from the steam steriliser immediately after the
sterilisation cycle is completed.
4
Use according to the manufacturer's Instructions for Use.
NOTE
Allow the sterilised items to cool to room temperature before using them
again.
6.5 Storage
4
Comply with all hygiene requirements when storing sterile items.
4
Store in a dry place protected from dust.
4
Note the labelling on the packaging.
4
Note the duration of storage.