Instructions for use EXPERTsurg LUX – 1.008.3500
5 Operation | 5.7 Activating the one-touch calibration
43 / 68
4
To terminate the rinsing function, press the back key or the pump key.
5.6.2 Programme step Rinsing function
The rinsing function serves to feed coolant and to startup the illumination on
the handpiece. The motor is not activated during this process.
The rinsing function can be defined as an activity in the course of the pro-
gramme.
4
Defining and executing programme steps.
Also refer to:
5.4.2 Exemplary programme step sequences, Page 35 (Example 2)
5.7 Activating the one-touch calibration
The one-touch calibration automatically compensates for torque deviations of
the motor that may be caused, e.g., by aging processes. With the handpiece
attached, the unit detects if the handpiece runs sluggish or is defective. The
one-touch calibration thus provides for a more accurate torque on the contra-
angle handpiece.
NOTE
The handpiece must be attached for calibration.
One-touch calibration should be carried out only with KaVo surgical hand-
pieces with a transmission ratio of 16:1, 20:1 or 27:1.
The one-touch calibration cannot be run with third-party handpieces or hand-
pieces with different transmission ratios.
The calibration must be repeated whenever the handpiece is changed.
4
Press the arrow keys until the one-touch calibration is highlighted.
4
Press the plus or minus keys to start the one-touch calibration.
ð
The display shows "Press foot control".
Summary of Contents for 1.008.3500
Page 1: ...Instructions for use EXPERTsurg LUX 1 008 3500...
Page 67: ......
Page 68: ...1 010 3558 bd 20220901 11 en...