Instructions for use EXPERTsurg LUX – 1.008.3500
6 Reprocessing steps in accordance with ISO 17664 | 6.2 Automated reprocessing
50 / 68
6.2 Automated reprocessing
6.2.1 Automated cleaning and disinfection
CAUTION
Damage by liquids.
Faults on electrical components.
4
Do not subject the device to automated reprocessing.
The following parts of the unit are released for automated cleaning and disinfec-
tion:
▪ Handpiece holder
▪ Motor and motor cable
NOTE
Please comply with the corresponding Instructions for use when you repro-
cess the motor and motor cable.
The reprocessing steps for the handpiece holder are as follows:
KaVo recommends washer disinfectors in accordance with EN ISO 15883-1,
which are operated using alkaline cleaning agents.
The validation was performed in a Miele washer disinfector using the
"VARIO-TD" programme and the "neodisher MediClean" cleaner supplied by Dr.
Weigert.
In addition, KaVo recommends the use of a neutraliser and a rinsing agent.
4
For programme settings and the adaptation options to be used, please refer
to the Instructions for Use of the washer disinfector.
6.2.2 Automated drying
The drying procedure is usually part of the cleaning programme of the washer
disinfector.
NOTE
Please comply with the instructions for use of the washer disinfector.
6.2.3 Service, inspection and testing after preparation
NOTE
It is essential to comply with the hygiene requirements (sterility) during the
test after reprocessing. If any fracturing or obvious surface changes are vis-
ible, the parts need to be checked by the Service.
Check for cleanliness, intactness, servicing, and repair as described in the fol-
lowing:
4
Check the adjustable functions of the unit and the motor function.
4
Check the hose pump for sufficient coolant flow rate.
4
Check the control commands on the foot control.
6.3 Packaging
NOTE
The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!