Instructions for use EXPERTsurg LUX – 1.008.3500
13 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 13.1 Operating envir-
onment and EMC warning notes
63 / 68
13 Information concerning the electromagnetic
compatibility according to EN IEC 60601-1-2
13.1 Operating environment and EMC warning notes
This product is neither life-sustaining nor coupled to the patient. It is suitable, in
terms of EMC, for operation both in domestic healthcare and in facilities used
for medical purposes except in the vicinity of active facilities of HF surgery
devices or in rooms/areas, in which EMC interference of high-intensity may oc-
cur.
The customer or user has to ensure that the device is set up and used in an en-
vironment of the specified type and/or in accordance with the specifications of
the manufacturers.
This device uses HF energy for its internal functions exclusively. Therefore, the
HF emission of the device is very low and interference with adjacent electronic
devices is unlikely.
WARNING
Use of other devices next to this product
Faulty operation
4
The use of other devices in the immediate vicinity of this product or placed
on top of this product should be avoided as this may lead to faulty operation
of the unit. However, if this kind of use were necessary regardless, this
device and the other devices should be monitored closely in order to make
sure that they work properly.
WARNING
Non-approved accessories
Electromagnetic interferences
4
The use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this device can lead to increased electro-
magnetic emissions or reduced electromagnetic immunity of the device and
faulty operation thereof.
WARNING
Portable HF communication devices
Reduced performance features
4
Portable HF communication devices (including their accessories, such as,
e.g., antenna cables and external antennas) should be used no closer than
30 cm (or 12 inch) from the parts and cables of the product designated by
the manufacturer. Non-compliance can lead to a reduction of the perform-
ance features of the device.
13.2 Results of the electromagnetic tests
Requirement
Class / test level
Electromagnetic emissions
DIN EN 55011
VDE 0875-11 / 04.2011
Conducted emitted interference
[150 kHz – 30 MHz]
Class B
Radiated emitted interference
Class B