Instructions for use EXPERTsurg LUX – 1.008.3500
5 Operation | 5.2 Make the device settings
30 / 68
The following device settings need to be made once during the initial startup of
the device and can be adapted according to need. The software version in use
can only be displayed, but not changed in the device settings.
View
Setting
Language
The LUX brightness determines the brightness of the
LEDs on the straight or contra-angle handpiece.
The brightness can be set in 4 steps ranging from "off"
to maximum brightness.
The LUX afterglow time determines how long the LEDs
of the straight or contra-angle handpiece afterglow
after each motor stop.
The afterglow time can be set in a range from 0 to 10
seconds.
Operating mode of the foot pedal
The motor can be triggered in either of two operating
modes:
▪ The motor starts at full speed once the foot pedal is
actuated
▪ The motor speed can be adjusted continuously up
to the maximum level using the foot pedal
Pump key operating mode (available from software
version 1.40)
The pump can be triggered in either of two operating
modes:
▪ Actuating the pump key switches the coolant flow
on or off
▪ Actuating the pump key gradually increases the
coolant flow up to its maximum value
The time can be set in units of hours, minutes and
seconds.
The date can be set in units of day, month and year.