Instructions for use EXPERTsurg LUX – 1.008.3500
5 Operation | 5.7 Activating the one-touch calibration
44 / 68
CAUTION
The motor starts at full speed.
Risk of injury.
4
Hold the motor firmly or put it in a safe holder during the calibration.
4
Press the foot control and hold it down until the display shows that "Calibra-
tion was successful".
4
If you release the foot control before the display shows that the calibration
was successful, press the foot control again until the display shows a mes-
sage indicating that the calibration was successful.
4
Press the back key to terminate the calibration and to return to the selection
of device settings.
If an unsuitable or defective handpiece was used in the calibration, the calibra-
tion is discontinued and the error message, "Measurement failed - Non-per-
missible current", is shown.
4
Press the back key to terminate the failed calibration.