Instructions for use EXPERTsurg LUX – 1.008.3500
10 Troubleshooting
56 / 68
10 Troubleshooting
WARNING
Repair WITHOUT using KaVo original spare parts.
Parts such as the cover can come loose and cause injury. Aspiration, swallowing
of parts and danger of suffocation.
4
Only use spare parts that comply with the specification for repair; original
KaVo spare parts comply with the specification.
NOTE
If a repair is done with NON-KaVo original spare parts, this may constitute a
product modification that leads to the loss of CE conformity. In the event of
damage, the responsibility is with the service company or the operator.
The introduction of a modified product in the market, in which the reasonable
suspicion exists to endanger the safety and health of patients or users, is pro-
hibited by Medical Device Law §4, Abs.1 No. 1 and therefore requires its own
conformity check.
NOTE
If a malfunction cannot be remedied using the description in the following
tables, have trained service personnel perform the repair. The technician's in-
structions are available to the service personnel for servicing and repair work.
In case of malfunction, the display names the malfunction directly or shows an
error number.
Malfunction
Cause
Remedy
Non-functional device.
The unit is switched off.
4
Switch-on the mains switch on the rear of
the unit.
Neither end of the power
cable is plugged in.
4
Plug in the power input cable.
Unknown.
4
Turn the unit off and on.
Blown fuse.
4
Notify the service.
No coolant in the handpiece.
No coolant flow pre-se-
lected.
Pump is off.
4
Pre-select coolant flow.
Also refer to:
5.5.3 Setting the coolant flow, Page 39
Hose fixation is not
plugged in.
4
Plug in the hose fixation.
Also refer to:
4.5 Connecting the coolant container and
hose set, Page 23
Bottle is empty.
4
Connect new bottle.
Hose clamp is closed.
4
Open the hose clamp.
Pump locking mechan-
ism is not closed.
4
Check and close, if needed, the closing
mechanism.
Hose is kinked.
4
Check hose and remove the kink, if any.
Glass bottle containing
the coolant is not ventil-
ated.
4
Open the cap on the ventilation valve of
the puncture needle.