Instructions for use EXPERTsurg LUX – 1.008.3500
2 Safety | 2.1 Explosion hazard
7 / 68
2 Safety
NOTE
All serious events occurring in relation to the product must be reported to the
manufacturer and the competent authority of the member state, in which the
user and/or patient resides.
The instructions for use are an integral part of the product and must be read
carefully prior to use and must be accessible at all times.
The device may only be used in accordance with the intended use, any other
type of use is not permitted.
2.1 Explosion hazard
Electrical sparks in the product can lead to explosion or fire.
4
Do not use the product in explosion hazard areas.
4
Do not operate the product in an oxygen-enriched atmosphere.
4
Do not use the product in the vicinity of flammable gases.
2.2 Infection hazard
Patients, users or third parties could be infected by contaminated medical
devices.
4
Take suitable personal protective measures.
4
Follow the instructions for use of the components.
4
Before initial startup and after each use, reprocess the product and ac-
cessories appropriately.
4
Carry out the reprocessing as described in the instructions for use. The pro-
cedure has been validated by the manufacturer.
4
If you deviate from this validated procedure, make sure that the repro-
cessing procedure is effective.
4
After each patient, dispose of the coolant reservoir and replace the hoses.
4
Reprocess the product and accessories appropriately before disposal.
2.3 Electrical shock
Use of the product in moist or electrically conductive environments can lead to
electrical shock and injury to patient, user and third parties.
4
Use the product in dry environments exclusively.
4
Use the product only in environments that are not electrically conductive.
4
Protect openings of the product from any ingress of liquids.
4
Do not place the product in a trough-like container
4
Check for and ensure the absence of leakage from the coolant containers
and lines.
4
If any liquid is detected on the device, disconnect the power cable from the
supply mains right away and do not touch the product.
4
Make sure that the surface of the product is absolutely dry before plugging
the power cable back in the socket.
4
After servicing, interventions on and repairs of the device and before re-use,
have the service personnel perform safety checks on the device.
Summary of Contents for 1.008.3500