APPENDIX 8-1
_______________________________________________________________________
Copyright © 2010 by Kaufman & Robinson, Inc., 1306 Blue Spruce Drive, Unit A, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com
8. APPENDIX
The auto controller is preset to match the product supplied with it so the user should
usually not need to enter the setup mode. The setup is provided in case there has
been a change to the original product.
8.1 Auto Controller Setup
1. Start with the controller off. Press and hold the Program select and the Ac-
quire Setpoints buttons while tuning on the power switch. Continue holding
the two buttons until the LED indicators on the front panel begin to
fl
ash.
Set the maximum
fl
ow for the mass
fl
ow controller that will be attached to
the four gas channels. Use the Gas Channel Select button to select each
channel that will be used, then use the Gas Adjust knob to dial in the cor-
rect maximum
fl
ow. When
fi
nished, press and release the Operating Mode
button.
2. At this Point, the display alternately
fl
ashes Lrn and on. This is the learn
mode for the auto controller. When the auto controller is being used to run
the discharge of an end-Hall ion source in auto mode, it will learn how to
reach the operating conditions the
fi
rst time a program is used. The next
time, the feedback will home in more quickly. Since this effects the starting
gas
fl
ow, it may be desirable to turn this off for some applications. Turn the
setpoint knob to select off if desired. When
fi
nished, press and release the
Operating Mode button. Note that the learn mode does not function when
operating the auto controller remotely using the remote setpoint option.
3. The display now reads bd3. This is the baud rate selection for RS232 com-
munication with the auto controller. Turn the Gas Adjust knob to select
the baud rates. 0=1200, 1=2400, 2=4800, 3=9600, 4=19200, 5=38400,
6=57600, 7=115200. After selecting the desired baud rate, press and re-
lease the Operating Mode button.
4. The display should now read L16. This is the led brightness for the indica-
tor LED’s (not the display) turn the Gas Adjust knob to obtain the desired
brightness from L2 to L64. When
fi
nished, press and release the Operating
Mode button.
5. The display should alternately
fl
ash HC4 and 10. The number 10 is the mini-
mum
fl
ow in sccm that the hollow cathode can be adjusted to when using
either the Auto Gas or Manual Gas Operating Mode. Use the gas adjust
knob to adjust the minimum
fl
ow to a value other than 10 if desired. When
fi
nished, press and release the Operating Mode button.