REMOTE/AUTOMATED CONTROL 6-13
_______________________________________________________________________
Copyright © 2010 by Kaufman & Robinson, Inc., 1306 Blue Spruce Drive, Unit A, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com
COM:1 – Place unit in ‘RS-232 Remote’ control.
If the unit is not in front panel remote and standby, the unit
issues the following response:
“Unit must be in STANDBY AND front panel REMOTE”.
COM:0 – Disable ‘RS232 remote’ control.
If the unit is not in ‘Standby’ when this command is issued, if
not the unit responds with:
“Unit must be in STANDBY”.
COM? – Query of remote control state.
.
If the unit is in ‘Verbose’, the response is “Enabled” or
“Disabled”.
If the unit is in ‘Terse’, the response is “1”, or “0”.
Example: COM:0<cr><lf> – Disable RS-232 remote control.
Example: COM:1<cr><lf> – Enable RS232 remote control.
Example: COM? <cr><lf> – Get present output state.
If COM is not ‘Enabled’, then the response to commands will be “Comm
Inactive” in both ‘Terse’ and ‘Verbose’ modes. The unit will respond to
query’s regardless of the COM state.
6.4.4.3 *IDN? - Unit Identi
fi
cation Query
The *IDN? command will return a unit identi
fi
cation string. The form of this
command follows the SCPI standard query form.
Command: *IDN?
Response:
The KRI Auto-Controller will return information in the following form:
<Manufacturer>,<Model>,<Front panel
fi
rmware version>, <Main
board
fi
rmware version>,<FPGA
fi
rmware version>. The form of the