brief; see the detailed descriptions above to understand the full functionality.
Table 6-5. Pin descriptions for the female DB-37 connector on the Auto Con-
troller.
Pin
Number
Description
1
Emission Current Output. 0-5 volts corresponds with 0 to the maximum
emission current in amps for the Emission Power Supply being used.
Referenced to pin 7 or 8.
2 No
Contact
3
Discharge Voltage Setpoint. 0-5 volts corresponds with 0 to the maxi-
mum output voltage for the Discharge Power Supply. Referenced to pin
7 or 8.
4
Emission current set point. 0-5 volts corresponds with 0 to the maxi-
mum emission current in amps for the Emission Power Supply being
used. Referenced to pin 7 or 8.
5
Aux 12 volts. Referenced to pin 7 or 8.
6
Emission Voltage Output. 0-5 volts corresponds with 0 to the maximum
output voltage for the Emission Power Supply. Referenced to pin 7 or 8.
7 Common.
8 Common.
9
Channel 1 Gas Flow Output. 0-5 volts corresponds with 0 to the maxi-
mum output in sccm for the channel 1 MFC. Referenced to pin 27.
10
Channel 3 Gas Flow Output. 0-5 volt corresponds with 0 to the maxi-
mum output in sccm for the channel 3 MFC. Referenced to pin 27.
11
Channel 1 Gas Setpoint. 0-5 volts corresponds with 0 to the maximum
output for the MFC connected to Channel 1. Referenced to pin 13.
12
Channel 3 Gas Setpoint. 0-5 volts corresponds with 0 to the maximum
output for the MFC connected to Channel 3. Referenced to pin 13.
13
Gas Setpoint Common.
14
Gas Channel 3 shut off. Connecting this pin to common pin 7 or 8 will
disable the gas
fl
ow to the MFC and deactivate the additional MFC relay
for channel 3, regardless of the input
fl
ow setting.