3.Display Description and Operation Procedure
20
3.5.3 INTERVAL
[Measurement Flow]
Configure INTERVAL, SAMPLE T and CYCLES, and repeat measuring.
Example: If the setting is as follows:
INTERVAL: 00:01:00
SAMPLE T.: 00:00:15
CYCLES: 20
A total of 20 measurements are performed at one minute intervals.
Note: Sampling time is 15 seconds and standby time is 45 seconds.
Measurement Flow
If the time period between two samplings is 10 seconds or longer, the pump will be turned OFF after
a sampling is finished. Then the pump will be turned ON again 10 seconds before the next sampling
starts.
[Measurement Settings]
You can set INTERVAL, SAMPLE T., CYCLES and
MANUAL STOP.
To set INTERVAL and SAMPLE T, tap the entry
field to display a numeric keypad. Enter “hours”,
“minutes” and “seconds”, and tap [OK].
Tap the entry field for CYCLES to display a
numeric keypad. Tap the value you want to
enter and tap [Enter].
Check “MANUAL STOP” to continue measuring until you stop measuring regardless of the CYCLES
setting.
Check “AUTO SAVE” to store the data automatically.
Check “AUTO PRINT” to print the data automatically.
Check “START DELAY” to use a delay time prior to a measurement.
Check “Set time” to set the time when you want to start a measurement.
Tap [OK] to save and activate the settings