3.Display Description and Operation Procedure
30
3.6.2 STATS
- When not using a MAP file
Tap [START] to display “CONFIRM” screen.
A measurement will start based on the
setting configured in
3.5 MODE
.
Tap [START] to start a measurement.
If “START DELAY” is NOT checked in
“MODE” setting:
Upon tapping [START], the pump will
begin running. After the pump has run
for
10
seconds
to
stabilize,
the
measurement will begin
If only “START DELAY” is checked in
“MODE” setting:
Upon tapping [START], the Delay Time
begins. During the last 10 seconds of
that Delay Time the pump will run to
stabilize, after which the measurement
will begin.
If “START DELAY” and “Set time”
are checked in “MODE” setting:
Upon tapping [START], “Input START
TIME” dialog box will be displayed. Enter
the
time
when
you
want
to
start
measuring and tap [OK]. During the last
10 seconds of the [Time left] time the
pump will run to stabilize, after which the
measurement will begin.
Tap [CANCEL] to stop measuring and go
back to the menu screen.
If you fail to press [OK] after entering
“START TIME”, and the entered START
TIME
arrives,
the
instrument
will
increase the start time in 10 minute
intervals until you press [OK].
<Input START TIME dialog box>