Important Safety Information
iv
¤
Do not move the main unit or the probe from a cold place to a warm place
quickly. It will cause condensation.
…… Even when used in an environment within the specified operating temperature and
humidity, a sudden temperature change may cause condensation. Condensation
generated on the sensor may cause inaccurate measurements. Condensation on metal
parts may cause rusting and lead to a malfunction.
¤
Do not touch the instrument when it is electrically charged.
……
Failure to observe the above may affect the measurement value and cause damage to
the instrument circuitry.
¤
Do not let the instrument suck in highly concentrated particles that exceed the
specification level.
……
Failure to observe the above may contaminate the instrument. As
a result the instrument may fail to obtain zero count.
In order to check the performance under highly concentrated
environments, make sure to attach the filter to the inlet.
If you accidentally let the instrument suck in highly concentrated
particles, attach the filter to the inlet and brush inside the
instrument while the pump is running.
¤
Do not dispose of the instrument as household waste.
…… Please note that the disposal of the instrument and the batteries should be in line with
your local or national legislation. For details, please contact your local distributor.