3.Display Description and Operation Procedure
24
3.5.5 STANDARD
[Measurement Flow]
You can perform a measurement which complies with 7 measurement standards as follows:
-
ISO14644-1 (International Standard)
Refer to ISO14644-1.
-
ISO 14644-1 SEQUENTIAL SAMPLING
(International Standard) Refer to ISO14644-1.
-
FEDERAL STANDARD (m)
(The US Standard) Refer to FEDERAL STANDARD 209E.
-
FEDERAL STANDARD (ft)
(The US Standard)
-
BRITISH STANDARD
(British Standard) Refer to BS-5295.
-
EC GMP
(European Standard) Refer to EC GMP.
-
GB/T 16292-1996
(Chinese Standard) Refer to GB/T 16292-1996.
You can set [SAMPLE T.], [CYCLES], and [SAMPLE POINT] to repeat measurements.
INTERVAL and SAMPLE T. are configured with the same value. Measurements are repeated
according to the CYCLES setting in each SAMPLE POINT.
After the all the measurements are finished, the results are evaluated according to the selected
standard.
Tap [STANDARD] to open the SELECT
STANDARD
window
to
select
a
measurement standard to be applied in
the STANDARD mode.
Select and set one of the above
measurement standards.
Tap [OK] to display the
MEASURE SETTING screen to
configure
measurement
settings.
Tap [CANCEL] to go back to
[MODE] screen.
[Measurement Settings]
The setting items for CLASS and items in the
EXTENSION tab varies depending on the
STANDARD settings you configured.
In the BASIC tab;
Configure the settings of CLASS, SCHEDULE,
ROOM, LOCATION, CYCLES and SAMPLE T.