6. Data Output
27
6.6 Analog Output (Optional)
1. Data Update Interval …… 1 second
2. Load Impedance………… 5KΩ and above
3
.
Output Voltage……………DC 0-1V
As For the analog output, you can select one output range from the
table below.
Output Range
Conversion Formula
(Voltage: V)
CO (C)
0 ~ 50 ppm
C = 50 ×V ppm
0 ~ 100 ppm
C = 100×V ppm
0 ~ 250 ppm
C = 250×V ppm
0 ~ 500 ppm
C = 500×V ppm
CO
2
(M)
0 ~ 500 ppm
M = 500×V ppm
0 ~ 1000 ppm
M = 1000×V ppm
0 ~ 2500 ppm
M = 2500×V ppm
0 ~ 5000 ppm
M = 5000×V ppm
Temperature
(T) 0 ~ 50
o
C
T = 50 ×V
o
C
0 ~ 100
o
C
T = 100×V
o
C
-20 ~ 30
o
C
T = 50 ×V – 20
o
C
-20 ~ 80
o
C
T = 100×V – 20
o
C
Temperature
(F)
32 ~ 122
o
F
F = 90 ×V + 32
o
F
32 ~ 212
o
F
F = 180×V + 32
o
F
-4 ~ 86
o
F
F = 90 ×V – 4
o
F
-4 ~ 176
o
F
F = 180×V – 4
o
F
Humidity (H)
0 ~ 50 %RH
H = 50 ×V %RH
0 ~ 100 %RH
H = 100×V %RH
It is linear output whose minimum value is 0V and maximum value is 1V.
Output data is output every 1 sec constantly.
How to Output Measurement Data (
Analog Output)
Explanation
Take data every 1 second,
and output the value every
1 second.
Analog Output
Terminal
0 1 2 3 4sec.
(Measurement Time)
Voltage