6. Data Output
25
6.5 Entering Commands from PC to Output Data
As for connecting your PC to the instrument,
refer to “6.4.1 Preparation for Digital Output” (Page 24).
-----Icon and its meaning-----
: Space
: Line Break or to Press [Enter] key
*
: To enter any number
*
Enter all commands in capitals.
6.5.1 Transmission of Raw Data (data measured every 1 sec)
Display
Procedure
Example:
When entering “D0005 ” in Measurement
Mode.
<To set number of data to be read>
Enter
“D
****
”
.
(Enter the number of data to be read in 4 digit
number.) After the command is received, “AD” will be returned. Then
the raw data which is displayed on the instrument’s screen every one
second will be output. The maximum number of data which can be set
is 9999. When to read more than 9999 data, send another command.
What will be output?
CO; CO2; Temperature; Humidity
<Receive Interrupt>
Enter
“N ”
. After the command is received, “AN” will be returned and
the reception will be interrupted
Display
Procedure
<To output measurement condition>
Enter
“S ”
. After the command is received, “AS” will be returned.
The display range of measurement items shown on the screen and
configured atmospheric pressure will be output.
What will be output?
CTH: CO measurement range; temperature measurement range;
configured atmosphere pressure
CO measurement range: 00:0~50PPM / 01:0~500PPM
Temperature measurement range: 00:0~60
o
C / 01:-20~60
o
C
Display
Procedure
<To output measurement unit >
Enter
“U ”
. After the command is received, “AU” will be returned.
Configured measurement unit at the time will be output.
What will be output?
CO Unit; CO
2
Unit; Temperature Unit; Humidity Unit;
Dew-point Temperature Unit; Wet-bulb Temperature Unit; Absolute
Humidity Unit; Humidity Ratio Unit; Percent Outdoor Air Unit
AN
AD
0.9;
576;
23.4;
63.4
0.8;
556;
23.4;
63.3
0.8;
534;
23.5;
63.2
0.9;
540;
23.5;
63.2
0.9;
561;
23.4;
63.3
AS
CTH;00;00;1013
AU
ppm;ppm;゚ C;%RH;゚ C;゚ C;g/m3;g/kg;%
Command
Function
D
****
To set how many data to read
N
Receive Interrupt
S
To output measurement condition
U
To output measurement unit
F
To set to output humidity related data
G
To cancel outputting humidity related data
P
To output stored number of the page
T
****
To output stored data
M
****
To output measurement condition
B
To output measurement condition of all the
pages