6 Operation
38
Simulation mode accesses the analog outputs and binary outputs directly.
When simulation mode has been selected, MANUAL mode is not possible!
In MANUAL mode the settings for "higher order controllers" are taken into
consideration.
6.8.1
MANUAL mode only via "higher order" controller functions
Select manual mode
✱
Set ADMINISTR. LEVEL/PARAMETER LEVEL/CONTROLLER/
CTRL.SPEC. FUNCT./MANUAL MODE "Locked, Coding or Switching.
Locked =
No Manual mode, control is via device.
Coding =
The outputs are active as long as the
or
key is pressed.
Switching = the outputs are active if the
or
key is pressed. If the
corresponding key is pressed again, the output becomes inactive
again.
Activate Manual mode
The device is in Display mode
✱
Press the
and
keys for less than 2 seconds.
The word MANUAL appears in the status line of the display.
Simulation mode
MANUAL mode
"Higher order"
controller
Binary outputs
Analog outputs
In the factory setting of the device the MANUAL mode parameter is locked and
can only be activated by the administrator!
This parameter must first be released for other users, see "Release level",
page 35.
EXIT